FREE WATER MAY BE PRESENT <1000 PPM

UnclassifiedLubrication

What This Code Means

Inaccurate water concentration measurements; Unacceptably large error in FluidScan results; Water measurements never exceed 6,000 ppm; Oil sample is not truly homogeneous despite appearing opaque; FluidScan reports traditional E2412 dissolved water result and displays alert message

Possible Causes

  • Hand-shaking is insufficient to obtain a homogeneous sample with uniform water droplet size
  • Sample prepared only with vigorous hand-shaking (Method B)
  • Total water calibration <1,000 ppm is not available for a fluid and the total water detected is <1,000 ppm
  • Total water calibration <1,000 ppm is not available for the fluid and total water detected is <1,000 ppm

Suggested Actions

  1. 1Measure a fresh oil sample immediately at the sampling site to leverage machine turbulence and shearing
  2. 2Prepare samples prior to analysis using a commercially available homogenizer